Chinese Professional Yuwell Ultrasonic Nebulizer - ACCUGENCE ® Multi-Monitoring System (PM 800) – e-Linkcare Detail:
Feature |
Specification |
Parameter |
Blood Glucose, Blood β-Ketone, and Blood Uric Acid |
Measurement Range |
Blood Glucose: 0.6 – 33.3 mmol/L (10 – 600 mg/dL) |
Blood β-Ketone: 0.0 – 8.0 mmol/L |
|
Uric Acid: 3.0 – 20.0 mg/dL (179 – 1190 μmol/L) |
|
Hemoglobin: 3.0-26.0 g/dL (1.9-16.1mmol/L) |
|
Hematocrit Range |
Blood Glucose and β-Ketone: 15% – 70 % |
Uric Acid: 25% – 60% |
|
Sample |
When testing β-Ketone, Uric Acid or Blood glucose with Glucose Dehydrogenase FAD-Dependent, use fresh capillary whole blood and venous Blood samples; |
When testing Blood glucose with Glucose Oxidase: use fresh capillary whole blood |
|
Minimum Sample Size |
Hemoglobin: 1.2 μL |
Blood Glucose: 0.7 μL |
|
Blood β-Ketone: 0.9 μL |
|
Blood Uric Acid: 1.0 μL |
|
Test Time |
Hemoglobin: 15 seconds |
Blood Glucose: 5 seconds |
|
Blood β-Ketone: 5 seconds |
|
Blood Uric Acid: 15 seconds |
|
Units of Measure |
Blood Glucose: The meter is preset to either millimole per liter (mmol/L) or milligrams per deciliter (mg/dL) depending on the standard of your country. |
Blood β-Ketone: The meter is preset to millimole per liter (mmol/L) |
|
Blood Uric Acid: The meter is preset to either micromole per liter (μmol/L) or milligrams per deciliter (mg/dL) depending on the standard of your country. |
|
Hemoglobin: the meter is preset to either to millimole per liter (mmol/L) or grams per deciliter (g/dL) depending on the standard of your country. | |
Memory |
Hemoglobin: 200 tests |
Blood Glucose: 500 tests (GOD + GDH) |
|
Blood β-Ketone: 100 tests |
|
Blood Uric Acid: 100 tests |
|
Automatic Shutoff |
2 minutes |
Meter Size |
86 mm × 52 mm × 18 mm |
On/Off Source |
Two CR 2032 3.0V coin cell batteries |
Battery Life |
Around 1000 tests |
Display Size |
32 mm × 40 mm |
Weight |
53 g (with battery installed) |
Operating Temperature |
Glucose and Ketone: 5 – 45 ºC (41 – 113ºF) |
Uric Acid: 10 – 40 ºC (50 – 104ºF) |
|
Operating Relative Humidity |
10 – 90% (non-condensing) |
Operating Altitude |
0 – 10000 feet (0 – 3048 meters) |
Product detail pictures:

Related Product Guide:
Our firm promises all consumers with the first-class products as well as most satisfying post-sale services. We warmly welcome our regular and new consumers to join us for Chinese Professional Yuwell Ultrasonic Nebulizer - ACCUGENCE ® Multi-Monitoring System (PM 800) – e-Linkcare , The product will supply to all over the world, such as: Netherlands, Karachi, Italy, We attained ISO9001 which provides solid foundation for our further development. Persisting in "High quality, Prompt Delivery, Competitive Price", we have established long-term cooperation with clients from both overseas and domestically and get new and old clients' high comments. It is our great honor to meet your demands. We are sincerely expecting your attention.